The
TS-300 is available in 2 versions. The TS-300LT is internally identical to
the smaller systems TS-150 and TS-140. Because the top plate is larger,
and therefore heavier, the maximum load is reduced to 120kg. The full
version TS-300 is a robust system having double the number of motors and
sensors, and able to support loads up to 300kg.

This is the largest of the TS-family and is ideally suited for
all applications which need more space on the top plate than is achievable
with the TS-140 / TS-150.
Drilled top plates with holes in metric (M6 x 25mm) or non-metric grid
allow mounting of various equipment on the isolated table top.
Frequency
Load range TS-300
Load range TS-300LT
Dimensions |
0,7-1000
Hz
0-300 kg
0-120 kg
600
x 800 x 120 mm |
|
|
The
TS-300 combines proven technical excellence with elegant and user
friendly design
 |
 |
 |
 |
 |
 |
|
| The
isolation of the TS-300 begins at 0,7 Hz,
increasing rapidly to 40dB beyond 10Hz
The
lack of any low frequency resonance means much better
performance than with most common passive air damping
approaches
The
inherent stiffness of the TS systems
imparts excellent directional+positional stability
The
outstanding performance of the TS-300 includes all six
translational and rotational vibration modes
The precise automatic height adjustment
mechanism stabilises the top plate at an optimised working
level even after major changes of load
A
remote control allows the isolation to be switched “ON” and “OFF” externally |
|
|
|
| |
|
|
|
|
|
|
 |
|
|
| Isolation: |
dynamic
0,7Hz to 1 kHz passive beyond 1 kHz |
| Transmissibility: |
see
attached curve Above 10 Hz trans- missibility
< 0,01 (-40dB) |
| Correction
Forces: |
TS-300 max.
+/- 8 N
TS-300LT max.
+/- 4 N
in any direction |
| Static
Compliance: |
TS-300: ~6 µm/N vertical
~15-20 µm/N
horizontal
TS-300LT: ~12 µm/N vertical
~30-40 µm/N
horizontal |
| Maximum
Load: |
TS-300 : 300
kg
TS-300LT: 120
kg |
| Size: |
600
x 800 x 120 mm |
| Weight: |
53
kg |
| Electrical |
|
| Safety
class: |
1 |
| Power
Consumption: |
TS-300: typically
20 W , maximum 100 W (140 VA)
TS-300LT: typically
10 W , maximum 50 W (70 VA) |
| Input
Voltage: |
115
- 230 V AC
+/- 10% 50 - 60 Hz |
| Overload: |
The
display indicates when the load is too heavy or severely
unbalanced |
| Monitor
signal: |
A
multiplexed signal for display on oscilloscope shows
vibration levels with and without isolation - for diagnostic
purposes only |
| Front
panel Display menu: |
Instrumental
status
accessible by scrolling:
- 3-axis vibration display
- height adjustment
- isolation ON / OFF
- overload
- transportation locking
- technical hot-line |
| Table
top |
|
| Size: |
600
x 800 mm |
| Material: |
Damped
Aluminium sandwich construction
Also available with tapped
M6 on 25mm
or customised |
|
|
information
please click:
Instruction Manual
TS-300
as
pdf
|
 |